This tool estimates the aerosol sample collection time required to reach a defined target mass load on an EMILIE™ chip. In the default mode it assumes a monodisperse aerosol of spherical particles with uniform diameter and density.
The collection time follows from:
t = Mtarget / (Q · N · CE(d, Q) · mp)
where Mtarget is the target chip load, Q the volumetric flow rate, N the number concentration, and mp = ρ · (π/6) · d³ the single-particle mass.
The size and flow dependent collection efficiency CE(d, Q) is described by an exponential fit calibrated from experimental data:
CE(d, Q) = CEmax(Q) · [1 − exp(−d / λ(Q))]
Calibrated at three flows: 0.3 and 0.6 L/min from Surdu et al. (2026); 1.0 L/min (maximum) from unpublished data provided by Gisinger. CEmax is interpolated linearly and λ log-linearly between adjacent calibration points. Particles below 10 nm are excluded as they are not reliably captured by the chip.